Plutonium Analysis by Alpha-Spectrometry
transparent spacer
 


Sample Receiving and Preparation
Sample Management Office
Sample Receiving
Quality Control


Analytical Processes
Thermal Ionization Mass Spectroscopy

TIMS Sample Preparation

Alpha-Spec Pu Sample Preparation - Separation

Alpha-Spec Pu Sample Preparation - Electroplating

Plutonium Analysis by Alpha-Spectrometry


Other Links
Project Documentation

Bioassay Homepage

 

 



The electroplated samples are analyzed and quantified by alpha-spectrometry. The samples are counted for 70,000 seconds on passivated ion-implanted junction silicon detectors, which provide an average energy resolution of 16 keV FWHM, allowing the activities of Pu-238 and Pu-239/240 to be determined based on chemical recovery of the Pu-242 tracer. The C-INC counting facility has dedicated 48 individual counting chambers to the LANL Bioassay Project. The nominal detection limit for Pu-238 and Pu-239/240 is less than 10 fCi/sample. This operation analyzes in excess of 3,000 samples per year for the LANL Bioassay Project.



 
transparent spacer transparent spacer transparent spacer transparent spacer transparent spacer


      C-ADI | C-AAC | C-PCS | C-ACS | C-INC | C-ACT | C-SIC | C-FM  

   

Operated by the University of California for the US Department of Energy
maida@lanl.gov | Copyright © 2000 UC | Disclaimer |

Last Modified: 8/12/02 NOTICE: Information from this server resides on a computer system funded by the U.S. Department of Energy. Anyone using this system consents to monitoring of this use by system or security personnel. For complete conditions of use see Disclaimer.

spacer spacer spacer spacer